Model Number |
Bio-A-2000 |
Biochemistry Analyser Model Bio-A-2000
The biochemistry Analyser Bio-A-2000 is a machines that process a large portion of the samples going into a hospital or private medical laboratory.
The Bio-A-2000 does many types of tests including enzyme levels (such as many of the liver function tests), ion levels (e.g. sodium and potassium), and other tell-tale chemicals (such as glucose, serum albumin, or creatinine).
The Bio-A-2000 comes with the following features and specifications:
- To be furnitured with 7 filters with the wavelength of 340, 405, 492, 510, 546, 578 and 630nm for detecting various items, and the Bio-A-2000 select working wavelength by itself. If you need filters other than these wavelengths, please specify it clearly as you order it.
- The machine can store 100 programs. You can program parameters either beforehand or at current time based on your needs.
- Automatically store the scale parameters of the first detection to convenient detections of the same item, which will be done directly sucking the sample to be detected.
- Quality control, automatically store, calculate and output quality control data.
- 128 x 240 LCD, user friendly menu system for convenient operation.
- Detect by means of end-point, kinetic, two-point, standard curve, and conc. & bichromatic.
- The Bio- A -2000, together with software developed by our company and computer, can build up a biochemistry workstation, which can complete comprehensive report, tracking report, statistic report, quality control for any item, and research and output function.
| Wavelength Range |
340nm to 700nm |
| Number of filters |
7 filters at 340, 405, 492, 510, 546, 578 and 630nm (others available if required) |
| Measurement Range |
0.001 to 3.000Abs |
| Precision |
± 0.001Abs |
| Repeatability |
≤0.005Abs |
| Linearity |
≤ ± 1% |
| Cross-contamination |
≤1% |
| Temperature Accuracy |
± 0.1 Degree centigrade |
| Temperature Control |
Built-in at 25, 30 and 37 ° C (user’s selectable) |
| Test Method |
end-point, kinetic, two-point , standard curve and simple/double wavelength test. |
| Programs |
Up to 100 programs can be built by user |
| Memory |
Storage capability of parameters & results with automatic quality control and calculations |
| Colorimetric Pool |
Titanium alloy |
| Reagent |
400 µl to 800 µl. Open System uses all kinds of reagent |
| Sample volume |
20 to 50µl |
| Display |
LCD Digital 128 x 240 mm large screen with display of Photometric absorbance, density, active unit, and normal parameters. User’s friendly menu provides step-by-step procedures |
| Printer |
Fitted-in |
|